 › Microprocessor Controlled..Intensive Care Incubator. › The temp controlled unit is operated by advanced micro computer technology. › The three digital display indication provide the Set Temperature, Air Temperature and Skin Temperature. › The safety features are self test and alarm system. Which will detect the different failures. › The removable heater and fan unit make it easy to clean and service. › In addition, the Skin/Air temperature display is larger then the other set temperature display. › Full visibility & accessibility of the baby is provided by the designed of acrylic hood. › A full complement of alarm provides safety convenience, with both audible and visual indicator. › Probe:- Activates if probe is open, short or unplugged seconds. › Fan:- Activates by fan failure or slow rotation. › Over Temperature:- Activates if the temperature sense 0.5 deg. Above the set temperature. › Low temperature:- Activates if the temperature sense 0.5 deg. Blow the set temperature. › Power Failure. › Programmable Set High/ Set Low temp & mute time. › Humidification is provided with the help of ultrasonic nebulizier which gives a constant humidification in the hood. › A big storage space is provided at the bottom of the unit to keep consumables etc. › The unit rests on four big castors with brakes in the front caster.
| Electrical Specification |
| Power |
200V/50Hz |
| Power consumption |
300Watts |
| Chasis Leakage |
500 Amp |
| Current Fuse |
5 Amp |
|