Technical Data Sheets
Usage
Sterilizer takes H2O2 as sterilizing agent and form plasmatic state of H2O2 by electromagnetic field under low temperature. It combines both gaseous and plasmatic H2O2 to make sterilization for the items in chamber and decomposes the residual H2O2 after sterilization.
It is widely used in CSSD or other medical institute for the sterilization of electric wire and cable, optical glasses and glass lens, power equipment and batteries, endoscopes, catheter, non-metallic and other wet or heat sensitive instruments.
| Technical data |
| 1 | Chamber Dimension / Volume | 750(D) × 450(W) × 400(H) mm / 135 L |
| 2 | Chamber Structure & Material | Rectangular, Aluminum alloy |
| 3 | Overall Dimension | 990(D) × 760(W) × 1600(H) mm |
| 4 | Door Type | Single/Double door, automatic vertical sliding, touch screen operation |
| 5 | Loading Method | Shelf loading, double layers |
| 6 | Installation Mode | Floor mounted |
| 7 | Programs | Standard cycle, Express cycle, Leak test |
| 8 | Sterilization Cycle Time | ≤50 min (Standard), ≤26 min (Express) |
| 9 | Sterilization Temperature | 50°C ±5°C |
| 10 | Sterilization Medium | H₂O₂ solution, 60% concentration |
| 11 | H₂O₂ Solutions injection mode | Cassette type,
12 capsules on each cartridge 2.2mI H2O2 per capsule. |
| 12 | H₂O₂ Consumption | 2 capsules per full cycle, 1 capsules per express cycle |
| 13 | Weight | 400 kg |
| 14 | Power Supply | Driving power: AC 380 V ± 10% 3 phases 50/60 Hz, 3.8Kw (Standard) |
| 15 | Display | 5.7” true color Touch screen display and operation |
| 16 | Data printer | Built-in thermal printer |
| 17 | Loading method | Loading shelves |
| 18 | Biological indicator | 56ºC, 48hº |