Features:
- Used with dialyzer and haemofilter for chronic renal failure treatment.
- Therapy Mode: HD IUF, On-line HDF
Online HDF:
- Dual-stage endotoxin filtration.
- Produce ultra-pure dialysate and substitution fluid online for priming and infusion reducing the use of saline to cut costs.
- Reduce complications and ensure the quality of treatment to improve patient survival significantly.
Design for Easy Operation
- Guided operations to set self-testing, parameters, treatments, cleaning, and disinfection.
- Clear management of permissions and roles (doctors, nurses, and engineers).
- Rotatable monitor with a 15-inch screen.
- Customized care and analysis enabled by multiple profiles.
Simple Maintenance
- Self-locking front panel for easy access and time savings.
- Modular, easy-to-install circuit board to simplify disassembly and maintenance.
Highly Compatible
- Compatible with all brands of disposables and a variety of dialysate and custom formulas, adaptable to all brands of concentrate and disinfectant.
- Heparin pump supports auto-identification and syringe matching.
Ultra Safe and Secure
- Utilization of ultrasonic and optical sensors to identify micro-bubbles and blood.
- Feedback control of volume balance for precise ultrafiltration.
- Feedback control of dialysate configuration to guarantee accurate ion concentration.
Cost-Effective
- Central concentrate supply to prevent secondary pollution.
- Online priming and reinfusion.
| Technical Specifications |
| Basic Information |
| Dimensions (W*D*H) |
465mm*785mm*1515mm |
| Net Weight |
87kg |
| Screen Size |
15 inches |
| Heparin Pump |
| Control Range |
0-10ml/h |
| Accuracy |
±0.2ml/h or ±5% of whichever is greater |
| Screen Size |
Bolus Volume Range |
|
| Operating Conditions |
| Voltage |
100-240V, 50/60Hz |
| Battery |
Supplies backup power for the blood pump and blood circuit monitoring system for not less than 40 minutes. |
| Water Inlet Temperature |
5-30°C |
| Water Inlet Pressure |
0.5bar-8.0bar |
| Communication Interface |
RJ45 RS232 USB |
|
| Ultrafiltration Control |
| Balance chamber ultrafiltration control system, real-time monitoring of leakage UF Rate |
0-4000ml/h |
| Accuracy |
±1% or ±30ml/h of whichever is greater |
|
| Blood Rate & Dialysate Control |
| Blood Flow |
0, 30ml/min-600ml/min
Accuracy: ±10ml/min or ±10% of whichever is greater
|
| Dialysate Flow |
100ml/min 1000ml/min Accuracy: ±5% |
| Conductivity Control |
12.5mS/cm-16.0mS/cm
Accuracy: ±0.1mS/cm
|
| Dialysate Temperature |
33°C-40°C
Accuracy: ±0.5°C
|
| Substitution Fluid Flow |
30ml/min-350ml/min
Accuracy: ±10ml/min or ±10% of whichever is greater
|
|
| Automatic Disinfection |
| Daily disinfection mode can be preset, and all disinfection conditions can be recorded, ensuring safety and ease of operation. |
| Decalcification |
Citric Acid |
| Heat Disinfection |
Hot Water or Citric Acid (≥93°C)
|
| Chemical Disinfection |
Sodium Hypochlorite |
|
| Monitoring System |
| VP Monitoring |
-60-520 mmHg Accuracy: ±10 mmHg |
| AP Monitoring |
-300-280 mmHg Accuracy: ±5 mmHg |
| TMP Monitoring |
-60-520 mmHg Accuracy: ±10 mmHg |
| Blood Leakage Detection |
0.35 ml/min blood at HCT of 32% |
| Air Detection single bubble |
>0.02 ml or continuous tiny bubbles of size more than 0.0003 ml |
| Blood Recognition |
Colour recognition, light monitor recognition |
|
| Non-invasive blood pressure monitor (optional) |
| Display range |
0mmHg~300mmHg |
| Monitoring range (Adult) |
40mmHg~245mmHg |