Usage :
Full support perspective, gastrointestinal spot film, gi (bariummeal, barium enema), orthopedic photography, pediatrics photography, chest, urinary system angiography, peripheral angiography operations, gynecological
photography (hsg) and many other checks, the real machine can realize all these function.
Features :
- High Performance and Intelligent SID electrically stetched to 1.8m Precise digital spot imaging Automatic stiching function-accurate and precise(optional)
- Cloud Image Interconnection DAP-dose visualized, safety assured(optional) Pulsed fluoroscopy, low dose and sharp imqges(optional)
- Stable and Reliable Image Chain Self-developped utrahigh frequency and high voltage generator Imported high quality X-ray tube Dynamic flat panel detector Removed dual grid
Project |
Content |
Technical parameter |
Power supply |
Voltage |
380v ± 38v |
Frequency |
50hz ± 10hz |
Capacity |
>105kva |
Internal resistance |
<0.172 |
High voltage generator (FSQ65RF) |
Maximum output power |
65.5kw (104kv, 630ma) |
Output power |
63kw (100kv, 630ma 0.1s) |
Main inverter frequency |
500khz ± 20% |
Photographic tube voltage |
40kv~150kv, 1kv continuously adjustable for each gear |
Photographic tube current |
Adjustment of 10ma~800ma |
Photographic time |
Adjustment of 1ms~1000ms |
Photographic mas |
0.1~1000mas |
Fluoroscopy tube voltage |
40kv~125kv continuously adjustable |
Fluoroscopy tube current |
5ma~20ma (pulse fluoroscopy)
0.5ma~10ma (continuous fluoroscopy) |
High voltage transformer structure |
Chive power frequency |
Beam limiter (XSQ20) |
Beam limiter field lamp |
Led lamp bead, dc3.6v/5w |
Field exposure time |
It can be set as 5-45s and each step is 5s |
Brightness of visible light |
Center average illumination intensity: > 100lux |
Equivalent total filtration |
>1mmal |
X-ray tube assembly (E7252X) |
Target material |
Molybdenum target of rhenium tungsten alloy |
Field exposure time |
It can be set as 5-45s and each step is 5s |
Target angle |
12° |
Permanent filtration |
0.9mm al l 75kv |
Operating voltage |
40-150kv |
Temperature range of tubular vessels |
16 ~ 75c, in normal use |
Input energy |
Small focus 27kw (180hz, 0.1s) Large focus 75kw (180hz, 0.1s) |
Resistance heat capacity |
210kj (300khu) |
Maximum heat dissipation of anode |
475w (667khu/min) |
Module heat capacity |
900k.j (1250khu) |
Rotating anode rate |
9700/min (180hz ) |
Tube focus |
1.2/0.6 |
Diagnosis bed (ZDC20F) |
Range of bed up and down |
90 degrees ~ - 25 degrees |
The spot device moves longitudinally |
1000mm ± 20mm |
Horizontal movement range of bed |
220mm |
surface, Bearing capacity of bed plate |
200kg |
Tube extension (sid) |
1000mm~1800mm+20mm |
Rotary pedal |
360° optional |
Filter grid |
498.5 x 449mm 230l/inch 10:
1 focal length 100cm,
498.5 x 449mm 103l/ inch 10:
1 focal length 180cm manual switching |