The NET anesthesia machine are used by anesthesiologists, nurse anesthetists to support the administration of anesthesia. It is designed to provide an accurate and continuous supply of medical gases (such as oxygen and nitrous oxide), which are mixed with an accurate concentration of anaesthetic vapour (such as isoflurane), and then delivered to the patient at a safe pressure and flow.
Basic Features:
Optional : Patient monitor, Anesthetic gas monitor, AGSS, O2 /AIR blender, Hand held ETCO2 module.
Ventilation mode | IPPV (VCV)/A/C (SIPPV) SIMV / MANUAL / SIGH |
Frequency | 1bpm - 100bpm |
I/E | 0.2s - 6s 2:1 - 1:6 |
Tidal Volume VT | 0ml - 1500ml |
MV | >18L/min |
PEEP | 0cmH2O - 30cmH2O |
PTr | -20 2O - 20cmH2O |
Ftr, Pc, TIP | - |
Apnea Ventilation | OFF,5s - 60s |
SIGH | 1/100 - 5/100 |
Graphical display | 1.P-T(pressure – time) 2.F-T(flow – time) " |
ANESTHESIA AGENT | ADJUSTABLE SCOPE % (VOLUME PERCENTAGE) |
Halothane | 0 ~ 5 |
Enflurane | 0 ~ 5 |
Isoflurane | 0 ~ 5 |
MV | 0 ~ 5 |
Sevoflurane | 0cmH2O - 30cmH2O |
Basic Features:
Optional : Patient monitor, Anesthetic gas monitor, AGSS,Yoke, ACGO, ETCO2 module on screen.
Ventilation mode | IPPV (VCV)/A/C (SIPPV) SIMV / MANUAL / SIGH |
Frequency | 1bpm - 100bpm |
I/E | 0.2s - 6s 2:1 - 1:6 |
Tidal Volume VT | 0ml - 1500ml |
MV | >18L/min |
PEEP | 0cmH2O - 30cmH2O |
PTr | -20 2O - 20cmH2O |
Ftr, Pc, TIP | (OFF,0.5L/min - 30L/min), (-), (0-3s) |
Apnea Ventilation | OFF,5s - 60s |
SIGH | 1/100 - 5/100 |
Graphical display | 1.P-T(pressure – time) 2.F-T(flow – time) " |
Same as AN314 but with Electronic Valve & Ventilation modes includes pressure control and volume control limit.
Ventilation mode | IPPV(VCV) A/C(SIPPV) PCV / PSV / P / SIMV / V-SIMV / SIGH / MANUAL / PVC-VG |
Frequency | 1bpm - 100bpm |
I/E | 0s - 12s 4:1 - 1:8 |
Tidal Volume VT | 0ml - 1500ml |
MV | >18L/min(Adult) >10L/min(Children) |
PEEP | 0cmH2O - 30cmH2O |
PTr | -20 2O - 20cmH2O |
Ftr | OFF, 1L/min - 30L/min |
Pc | 5cmH2O - 60cmH2O |
TIP | 0 - 6s |
Apnea Ventilation | OFF, 5s - 60s |
SIGH | 1/100 - 5/100 |
Graphical display | 1. P-T(pressure – time) 2. F-T(flow – time) 3. P-V loop (pressure – volume loop) |
Basic Features :
OPTIONAL: Patient monitor, Anesthetic gas monitor, AGSS, Yoke, built-in Anesthetic module.
AN330A: Same as AN330 but with patient monitor.
AN330B: Same as AN330A but with Anesthetic gas monitor, AGSS, Yoke-3.
AN335: Same as AN330 but with 15 inches TFL LCD touch screen and blended gas warming & Bypass function.
Ventilation mode | IPPV (VCV) / A/C (SIPPV) PCV SIMV SIGH MANUAL |
Frequency | 1bpm - 100bpm |
I/E | 4:1 - 1:8(Adult) 2:1 - 1:8(Children) |
Tidal Volume VT | 0ml - 1500m |
MV | >18L/min(Adult) >10L/min(Children) |
PEEP | 0cmH2O - 30cmH2O |
PTr | -20cmH2 O - 0cmH2O |
Ftr | OFF, 0.5L/min - 30L/min |
Pc | 5cmH2O - 60cmH2O |
TIP | 0 - 6s |
Apnea Ventilation | OFF, 5s - 60s |
SIGH | 1/100 - 5/100 |
Graphical display | 1. P-T(pressure – time) 2. F-T(flow – time) 3. P-V loop (pressure – volume loop) 4. V-T(volume – time ) 5. ETCO2 -T (End-tidal Co2 - time) |
Basic Features:
Safety:
*AN300 Anesthesia Ventilator is outside the frame
Ventilation Modes | Flow meter | Flow meter | Tidal volume (Vt) | Frequency (Freq) | I/E |
---|---|---|---|---|---|
IPPV, Manual | O2(0.1-10L/min) N2O(0.1-10L/min) |
(0.1-10L/min) | 50 - 1600ml | 1-90bpm | 4:1 -6:1 |
Basic Features:
Safety:
*AN305 Anesthesia Ventilator is inside the frame
Ventilation Modes | Flow meter | Flow meter | Tidal volume (Vt) | Frequency (Freq) | I/E |
---|---|---|---|---|---|
IPPV, Manual | O2(0.1-10L/min) N2O(0.1-10L/min) |
(0.1-10L/min) | 50 - 1600ml | 1-90bpm | 4:1 -6:1 |
Categories
What our Clients are Saying
The organic reviews not hoax.
Meet Us At
Booth: F11
11th to 13th Jun. 2025
Miami Beach Convention Center, Miami Beach, Florida, USA
02nd to 04th Sep. 2025
Cape Town International Convention Centre, South Africa.
17th to 20th Nov. 2025
Düsseldorf, Germany
Let's profit together!
We are looking for distributors in your area.
Most Searched items
Made from natural latex rubber.